- Altitude
- 3000 m (9842 ft) maximum
- Analog Output Functional Mode
- Linear, PID
- Communication (optional)
- Analog:Five 0-20 mA or 4-20 mA analog outputs on each analog output moduleUp to two analog Input modules (0-20 mA or 4-20 mA). Each input module replaces a digital sensor input.Digital:Profibus DPV1 moduleModbus TCPProfinet IO moduleEthernet IP module
- Compatible Instruments
- Compatible Sensors and Analyzers / Software Version (Release Year)Amtax sc / V2.30 (2018) or higherA-ISE sc / V1.02 or higherAN-ISE sc / V1.08 (2013) or higherN-ISE sc / V1.02 or higherNitratax clear sc, Nitratax eco sc, Nitratax plus sc / V3.13 (2013) or higherNT3100sc/NT3200scPhosphax sc / V2.30 (2018) or higherPhosphax sc LR/MR/HR / V1.01 (2018) or higherTSS sc / V41.73 (2013) or higherSolitax sc / V2.20 (2013) or higherTU5300sc, TU5400sc / V1.34 (2017) or higherSS7 sc (in Bypass) / V1.06 (2006) or higherUltraturb sc / V3.06 (2017) or higher1720E / V2.10 (2006) or higherSonatax sc / V1.15 (2016) or higherCL17sc / V2.7 (2019) or higherCL10sc / V1.14 (2013) or higher9184sc, 9185sc, 9187sc* / V2.03 (2013) or higherUvas plus sc / V3.01 (2017) or higherLDO 2 sc* / V1.22 (2013) or higher3798sc* / V2.03 (2013) or higher3700sc + Inductive Conductive Digital Gateway 6120800 / V3.00 (2017) or higher3422sc, Analog 3400 + Contacting Cond. Digital Gateway 6120700 / V3.00 or higherpHD sc*, pHD-S sc / V3.10 (2016) or higher1200-S sc* / V2.04 (2013) or higher pHD analog + Digital Gateway 6120500 / V3.00 (2017) or higherRC and PC analog sensor + Digital Gateway for conventional analog pH and ORP sensors 6120600 / V3.00 (2017) or higher 8362sc* / V3.00 (2017) or higher*Hardware Version1 of instrument is not supported
- Compatible Network Technologies
- GSM 3G/4G (e.g. AT&T, T-Mobile, Rogers, Vodafone etc.)CDMA (e.g. Verizon)
- Compliance Certifications
- CE. ETL certified to UL and CSA safety standards (with all sensor types), FCC, ISED, KC, RCM, EAC, UKCA, SABS, C م (Morocco)
- Conduit Openings
- ½" NPT conduit
- Dimensions
- ½ DIN - 144 x 144 x 192 mm (5.7 x 5.7 x 7.6 in.)
- Display
- 3.5-inch TFT colour display with capacitive touchpad
- Enclosure Rating
- UL50E type 4X, IEC/EN 60529–IP 66, NEMA 250 type 4XMetal enclosure with a corrosion-resistant finish
- Indoor/Outdoor
- Outdoor installation in direct sunlight or UV radiation requires UV protection screen and/or sunroof
- Installation Category
- Category II
- Material Enclosures
- Polycarbonate, aluminum (powder coated), stainless steel
- Measurements
- Two device digital SC connectors
- Mounting
- Wall, Pole, or Panel Mounting
- Network Connectivity
- LAN: Two Ethernet connectors (10/100 Mbps)Cellular: External 4GWi-Fi
- Operating Temperature Range
- -20 to 60 °C (-4 to 140 °F) (8 W (AC)/9 W (DC) sensor load)-20 to 45 °C (-4 to 113 °F) (28 W (AC)/20 W (DC) sensor load) Linear derating between 45 and 60 °C (-1.33 W/°C)
- Outputs
- LAN + Modbus TCP
- Pollution Degree
- 4
- Power Requirements (Voltage)
- 100-240 VAC ±10%, 50/60 Hz; 1 A
- Power Supply
- With EU plug
- Protection Class
- I, connected to protective earth
- Relays
- Two relays (SPDT);Wire gauge: 0.75 to 1.5 mm² (18 to 16 AWG)AC controllerMaximum switching voltage: 100 - 240 VACMaximum switching current: 5 A Resistive/1 A Pilot DutyMaximum switching power: 1200 VA Resistive/360 VA Pilot DutyDC controllerMaximum switching voltage: 30 VAC or 42 VDCMaximum switching current: 4 A Resistive/1 A Pilot DutyMaximum switching power: 125 W Resistive/28 W Pilot Duty
- Sensor Input #1
- Digital
- Sensor Input #2
- Digital
- Software available
- RTC-N/DNRTC-N/DN(ORP)RTC-N/DN(DO)
- Storage Conditions
- -20 - 70 °C (-4 - 158 °F), 0 - 95% relative humidity, non-condensing
- USB Port
- Used for data download and software upload. The controller records approximately 20,000 data points for each connected sensor.
- Warranty
- 12 months
- Weight
- 3.7 lb (controller only, w/o modules)
- What's included?
- SC4500 Controller, RTC-N/DN, RTC-N/DN (ORP), RTC-N/DN(DO), with EU plug; includes mounting hardware
|