LXV325.99.30002				 
				
					TSS HT sc Turbidity and Suspended Solids Inline Probe, Stainless Steel 
				 
				
				
				
					 6882   
					
				 
				
				    
					
					    
						
					 
				 
				
			 
			
						
			    				
										Automatic Wiper 
					No 
										Cable Length 
					10 m (optional extension cables available) 
										Calibration Method 
					Turbidity Formazin or StablCal® Standard (at 800 NTU). Requires a calibration kit. Suspended Solids Sample specific, based on gravimetric analysis with a correction factor procedure. 
										Flow 
					Max. 3 m/s (the presence of air bubbles affects the measurement) 
										Length 
					332 mm 
										Maintenance Interval 
					1 h/month 
										Material 
					Optics Carrier and Sleeve: stainless steel 1.4460 / stainless steel 1.4571 
										Max. cable length Sensor-to-Controller 
					100 
										Measuring Principle 
					Turbidity: 2-channel 90° scattered light measurement in accordance with DIN EN ISO 7027TSS: Modified absorption measurementEight-channel multiple-angle measurement, wavelength = 860 nm 
										Model 
					TSS HT sc 
										Operating Temperature Range 
					0 - 90 °C 
										Parameter 
					Turbidity, Suspended Solids 
										Pressure Range 
					< 10 bar or < 100 m 
										Range 
					Turbidity (TRB): 0.001 - 9999 NTUSolids (TSS): 0.001 - 500 g/L 
										Repeatability 
					TSS content: < 4 %Turbidity: < 3 % 
										Response Time T90 
					1 - 300 s adjustable 
										Special Notes 
					Installation note:Distance sensor–wall > 10 cm (TSS) 
										Warranty 
					12 months 
										Weight 
					approx. 1.6 kg 
										What's included? 
					Turbidity & Suspended Solids sensor, user manual 
									 
							 
			 
						
				
					 
				
					LXV328.99.30002				 
				
					TSS EX1 sc Turbidity and Suspended Solids Inline Probe, Stainless Steel 
				 
				
				
				
					 7270   
					
				 
				
				    
					
					    
						
					 
				 
				
			 
			
						
			    				
										Automatic Wiper 
					No 
										Cable Length 
					10 m (optional extension cables available) 
										Calibration Method 
					Turbidity Formazin or StablCal® Standard (at 800 NTU). Requires a calibration kit. Suspended Solids Sample specific, based on gravimetric analysis with a correction factor procedure. 
										Flow 
					Max. 3 m/s (the presence of air bubbles affects the measurement) 
										Length 
					488 mm 
										Maintenance Interval 
					1 h/month 
										Material 
					Optics Carrier and Sleeve: stainless steel 1.4460 / stainless steel 1.4404 
										Max. cable length Sensor-to-Controller 
					100 
										Measuring Principle 
					Turbidity: 2-channel 90° scattered light measurement in accordance with DIN/EN 27027/ISO7027TSS: Modified absorption measurementEight-channel multiple-angle measurement, wavelength = 860 nm 
										Model 
					TSS EX1 sc 
										Operating Temperature Range 
					-10 - 50 °C 
										Parameter 
					Turbidity, Suspended Solids 
										Pressure Range 
					10 bar or  100 m 
										Range 
					Turbidity (TRB): 0.001 - 9999 NTUSolids (TSS): 0.001 - 500 g/L 
										Repeatability 
					TSS content: < 4 %Turbidity: < 3 % 
										Response Time T90 
					1 - 300 s adjustable 
										Special Notes 
					Installation note:Distance sensor–wall > 10 cm (TSS) 
										Warranty 
					12 months 
										Weight 
					approx. 2.7 kg 
										What's included? 
					Turbidity & Suspended Solids sensor, user manual 
									 
							 
			 
						
				
					 
				
					LXV327.99.10002				 
				
					TSS XL sc Turbidity and Suspended Solids Inline Probe, Stainless Steel 
				 
				
				
				
					 6882   
					
				 
				
				    
					
					    
						
					 
				 
				
			 
			
						
			    				
										Automatic Wiper 
					No 
										Cable Length 
					10 m (optional extension cables available) 
										Calibration Method 
					Turbidity (TRB): Factory calibratedSolids (TS): To be calibrated by customer on siteZero point: Permanently calibrated in the factory 
										Flow 
					Max. 3 m/s (the presence of air bubbles affects the measurement) 
										Length 
					TSS XL sc: 232 mm 
										Maintenance Interval 
					1 h/month 
										Material 
					Gasket: FKM; Wiper: PA (GF), TPV 
										Measuring Principle 
					Combined multiple beam alternating light method with infrared diode system and beam focusingTurbidity: 2-channel 90° scattered light measurement in accordance with DIN/EN 27027/ISO 7027, wavelength = 860 nmSolids: Modified absorption measurement; eight-channel multiple-angle measurement, wavelength = 860 nm 
										Model 
					TSS XL sc 
										Operating Temperature Range 
					0 - 80 °C 
										Parameter 
					Turbidity, Suspended Solids 
										Pressure Range 
					16 bar or  160 m 
										Range 
					Turbidity (TRB): 0.001 - 9999 NTUSolids (TSS): 0.001 - 500 g/L 
										Repeatability 
					TSS content: < 4 %Turbidity: < 3 % 
										Response Time T90 
					1 - 300 s adjustable 
										Special Notes 
					Installation note:Distance sensor–wall > 50 cm (Turbidity)> 10 cm (TSS) 
										Warranty 
					12 months 
										Weight 
					approx. 1.5 kg 
										What's included? 
					Turbidity & Suspended Solids sensor, user manual 
									 
							 
			 
						
				
					 
				
					LXV326.99.10002				 
				
					TSS Vari sc Turbidity and Suspended Solids Inline probe, Stainless Steel 
				 
				
				
				
					 6882   
					
				 
				
				    
					
					    
						
					 
				 
				
			 
			
						
			    				
										Automatic Wiper 
					No 
										Cable Length 
					10 m (optional extension cables available) 
										Calibration Method 
					Turbidity (TRB): Factory calibratedSolids (TS): To be calibrated by customer on siteZero point: Permanently calibrated in the factory 
										Flow 
					Max. 3 m/s (the presence of air bubbles affects the measurement) 
										Length 
					TSS VARI sc 232 mm 
										Maintenance Interval 
					1 h/month 
										Material 
					Gasket: FKM; Wiper: PA (GF), TPV 
										Measuring Principle 
					Combined multiple beam alternating light method with infrared diode system and beam focusingTurbidity: 2-channel 90° scattered light measurement in accordance with DIN/EN 27027/ISO 7027, wavelength = 860 nmSolids: Modified absorption measurement; eight-channel multiple-angle measurement, wavelength = 860 nm 
										Model 
					TSS Vari sc 
										Operating Temperature Range 
					0 - 80 °C 
										Parameter 
					Turbidity, Suspended Solids 
										Pressure Range 
					<16 bar or <160 m 
										Range 
					Turbidity (TRB): 0.001 - 9999 NTUSolids (TSS): 0.001 - 500 g/L 
										Repeatability 
					TSS content: < 4 %Turbidity: < 3 % 
										Response Time T90 
					1 - 300 s adjustable 
										Special Notes 
					Installation note:Distance sensor–wall > 50 cm (Turbidity)> 10 cm (TSS) 
										Weight 
					approx. 1.5 kg 
										What's included? 
					Turbidity & Suspended Solids sensor, user manual 
									 
							 
			 
						
				
					 
				
					LXV440.99.00002				 
				
					AN-ISE sc Low cost ISE combination Ammonium and Nitrate probe (immersion) with RFID, 10 m cable 
				 
				
				
				
					 11091   
					
				 
				
				    
					
					    
						
					 
				 
				
			 
			
						
			    				
										Cable Length 
					10 m (32.8 ft) fixed cable at sensor, available up to 100 m 
										Calibration Method 
					With Cartrical plus technology: automatic import of factory calibration data from cartridge to probe by RFID; 1 and 2 point matrix correction 
										Flow 
					< 4 m/s max. velocity 
										Length 
					320 mm 
										Material 
					Cartridge: stainless steel (1.4571), PVC, POM, ABS, NBRCleaning unit: TPE, PUR, stainless steel (1.4571)Sensor: stainless steel (1.4571), ASA + PC, silicon, PVC and PU 
										Operating Temperature Range 
					Air: -20 - 45 °C (-4 - 113 °F) 
										Parameter 
					NH4-N and NO3-N 
										Protection Class 
					IP68 
										Range 
					0 - 1000 mg/L NH4-N0 - 1000 mg/L NO3-N0 - 1000 mg/L K+0 - 1000 mg/L Cl- 
										Repeatability 
					± 5 % of the measured value (with standard solution)+ 0.2 mg/L  NH4-N and  NO3-N 
										Storage Conditions 
					5 to 40 °C (41 to 104 °F) sensor cartridge 
										Warranty 
					12 months 
										Weight 
					2.38 kg 
										What's included? 
					Ammonium & Nitrate probe, sensor cartridge, user manual 
									 
							 
			 
						
				
					 
				
					HACH9070800				 
				
					 USB Cable 
				 
				
				
				
					 28   
					
				 
				
				    
					
					    
						
					 
				 
				
			 
			
						
			    			 
			 
						
				
					 
				
					HACH3422A1A				 
				
					Contacting Conductivity Sensor, Low Conductivity (k=0.05), with ½ 
				 
				
				
				
					 682   
					
				 
				
				    
					
					    
						
					 
				 
				
			 
			
						
			    				
										Cable Length 
					6 m (20 ft) 
										Flow 
					0 - 3 m/s maximum, fully immersed 
										Length 
					193.5 mm 
										Material 
					Compression Fitting: 1/2" NPT PVDF 
										Operating Temperature Range 
					-20 - 200 °C 
										Pressure Range 
					1.7 bar at 150 °C 
										Range 
					0 - 100 µS/cm 
										Repeatability 
					±0.5 % of reading 
										Response Time T90 
					30 s 
										Warranty 
					12 months 
										What's included? 
					Includes: sensor with 6 m (20 ft) cable and manual 
									 
							 
			 
						
				
					 
				
					HACH3433B8A				 
				
					General Purpose Contacting Conductivity Sensor, PPS Body, Low Conductivity (k=0.5) 
				 
				
				
				
					 564   
					
				 
				
				    
					
					    
						
					 
				 
				
			 
			
						
			    				
										Cable Length 
					20 ft 
										Flow 
					0 - 3 m/s 
										Length 
					134.6 mm 
										Operating Temperature Range 
					-20 - 200 °C 
										Pressure Range 
					6.8 bar at 150 °C (when used with hardware, a lower rated mounting hardware or piping material maylimit the temperatureand pressure ratings) 
										Range 
					0 - 1000 µS/cm 
										Repeatability 
					±0.5 % of reading 
										Response Time T90 
					< 30 s 
										Warranty 
					12 months 
										What's included? 
					Includes: sensor with 6 m (20 ft) cable and manual 
									 
							 
			 
						
				
					 
				
					HACH3422B3A				 
				
					Contacting Conductivity Sensor for Low Conductivity, K=0.5, 3/4 
				 
				
				
				
					 682   
					
				 
				
				    
					
					    
						
					 
				 
				
			 
			
						
			    				
										Cable Length 
					6 m (20 ft) 
										Flow 
					0 - 3 m/s maximum, fully immersed 
										Length 
					193.5 mm 
										Material 
					Compression Fitting: 3/4" NPT PVDF 
										Operating Temperature Range 
					-20 - 200 °C 
										Pressure Range 
					1.7 bar at 150 °C 
										Range 
					0 - 1000 µS/cm 
										Repeatability 
					±0.5 % of reading 
										Response Time T90 
					30 s 
										Warranty 
					12 months 
										What's included? 
					Includes: sensor with 6 m cable and manual 
									 
							 
			 
						
				
					 
				
					HACH3422C3A				 
				
					Contacting Conductivity Sensor, Mid-Range Conductivity (k=1.0), with ¾ 
				 
				
				
				
					 682   
					
				 
				
				    
					
					    
						
					 
				 
				
			 
			
						
			    				
										Cable Length 
					6 m (20 ft) 
										Flow 
					0 - 3 m/s maximum, fully immersed 
										Length 
					193.5 mm 
										Material 
					Compression Fitting: ¾" NPT PVDF 
										Operating Temperature Range 
					-20 - 200 °C 
										Pressure Range 
					1.7 bar at 150 °C 
										Range 
					0 - 2000 µS/cm 
										Repeatability 
					±0.5 % of reading 
										Response Time T90 
					30 s 
										Warranty 
					12 months 
										What's included? 
					Includes: sensor with 6 m (20 ft) cable and manual